J Korean Soc Neonatol. Search

CLOSE


Neonatal Med > Volume 17(2); 2010 > Article
Journal of the Korean Society of Neonatology 2010;17(2):239-244.
DOI: https://doi.org/10.5385/jksn.2010.17.2.239    Published online November 15, 2010.
A Comparative Study of Two Different Heel Lancet Devices for Blood Collection in Preterm Infants.
Hyo Bin Lim, Mi Joo Rhu, Ji Mi Jung, Ga Won Jeon, Jong Beom Sin
1Department of Pediatrics, Good Gangan Hospital, Korea.
2Department of Pediatrics, Inje University College of Medicine, Busan Paik Hospital, Busan, Korea. pedsin@inje.ac.kr
Abstract
PURPOSE
To evaluate two different heel lancet device in terms of pain response and success of the procedure in the preterm infants undergoing heel puncture.
METHODS
100 preterm infants undergoing capillary blood gas analysis or capillary bilirubin monitoring underwent heel puncture, were randomly allocated to blood sampling from the heel with either a conventional manual lancet or an automatic incision device. Primary outcome measures included the Premature Infants Pain Profile (PIPP) score, total duration of procedure, number of heel puncture and number of bruise. The pain response was evaluated using PIPP score and the effectiveness was evaluated using three criteria: total duration of blood sampling, number of puncture, bruising of the heel or ankle. Statistical analysis was performed using the SPSS ver. 13.0 program. Difference between the groups were analysed with t test (continuous variables) and the Chi square test or Fisher test (categorical variables).
RESULTS
The mean PIPP score was 4.91 for the automatic lancet group compared with 5.84 for the conventional manual lancet group (P=0.0255).The number of pain scores above 7 during blood collection did not differ between two groups (P=0.2167). The procedure took less time to perform in the automatic lancet group (mean, 30.69 seconds) than in the conventional lancet group (mean, 48.92 seconds) (P<0.0001).
CONCLUSION
This study demonstrated that the automatic lancet device causes less pain and a shorter procedure time than the conventional manual lancet in preterm infants undergoing heel puncture. On the basis of these results the automatic lancet device is very useful method for blood collection in preterm infants by heel puncture.
Key Words: Pain, Heel puncture, Automatic lancet device
TOOLS
Share :
Facebook Twitter Linked In Google+ Line it
METRICS Graph View
  • 3 Crossref
  •    
  • 1,962 View
  • 1 Download


ABOUT
ARTICLE CATEGORY

Browse all articles >

BROWSE ARTICLES
AUTHOR INFORMATION
Editorial Office
Rm.1207, King's garden 3 Block, 34, Sajik-ro 8-gil, Jongno-gu, Seoul 03174, Korea
Tel: +82-2-730-1993    Fax: +82-2-730-1994    E-mail: neonate2002@hanmail.net                

Copyright © 2018 by The Korean Society of Neonatology. All rights reserved.

Developed in M2community

Close layer
prev next