Using the High Frequency Ventilation during Neonatal Transport.
Seung Hyun Lee, Moon Hee Chae, Hye Jung Cho, Eell Ryoo, So Yeon Shim, Dong Woo Son
Department of Pediatrics, Graduate School of Medicine, Gachon University, Incheon, Korea. sondw@gilhospital.com
Abstract
PURPOSE
To present and evaluate a system of high-frequency oscillatory ventilator (HFOV) during intra-/inter-hospital neonate transport.
METHODS
The system includes a charged HFOV (SOPHIE, Fritz Stephan GmbH, Dusseldorf, Germany), an incubator, and E-oxygen/air-cylinders with connections to the HFOV. The test lung was evaluated at the high and medium ventilator settings used for infants to determine the operating time of HFOV. The time required to exhaust the gas supply was checked, and the HFOV was operated until the low-battery alarm sounded to determine the operating time of the batteries.
RESULTS
The batteries provided electrical power for at least 60 mins, and the oxygen and air-cylinders lasted at least 20 mins. The system has been used frequently for the intra-hospital transport, from delivery rooms to ICU and from ICU for surgery. The system has been used twice for the inter-hospital transport of infants with bronchopulmonary dysplasia and pulmonary hypertension to another hospital 45 km away (one hour distance). In one case, the ambulance's electrical power supply failed, causing the system failure during the last 5 mins of transport. However, with the complete check and simulation of the system and the ambulance bulk oxygen/electric supply, the second patient was transported successfully in stable condition.
CONCLUSION
The system was useful for intra-/inter-hospital transport of the neonates on HFOV. For the transport time of 60 mins, fully charged HFOV, 2 E-oxygen-cylinders, and 3 E-air-cylinders seemed to be sufficient. H-oxygen-cylinder and ambulance electrical power supply should also be provided for safe and efficient transport between hospitals.
Key Words: Bronchopulmonary dysplasia; Electricity; Equipment design; High-frequency oscillation ventilation; Newborn infant; Transportation of patients/methods; Mechanical ventilators
TOOLS
Share :
Facebook Twitter Linked In Google+ Line it
METRICS Graph View
  • 3 Crossref
  •    
  • 4,312 View
  • 30 Download


ABOUT
ARTICLE CATEGORY

Browse all articles >

BROWSE ARTICLES
AUTHOR INFORMATION
Editorial Office
34, Sajik-ro 8–gil(King’s Gargen 3 Block 1207), Jongno-gu, Seoul 03174, Republic of Korea
Tel: +82-2-730-1993    Fax: +82-2-730-1994    E-mail: neonate2002@naver.com                

Copyright © 2024 by The Korean Society of Neonatology.

Developed in M2PI

Close layer
prev next