Journal of the Korean Society of Neonatology 2000;7(1):33-38.
Published online January 1, 2001.
The Ventilatory Effect of Hybrid Ventilation in Rabbits.
Kook Hyun Lee, Hong Ko, Yong Seok Oh, Sang Chul Lee, Sang Whan Do, Yong Lak Kim
Department of Anesthesiology, College of Medicine, Seoul National University, Seoul, Korea.
Abstract
PURPOSE
Intratracheal pulmonary ventilation (ITPV) is developed to decrease dead space ventilation. A reverse thrust catheter (RTC) is introduced into an endotracheal tube through an adapter. Bias gas through the RTC exits from the catheter tip. The flow of gas is redirected outward away from the lung. Gas is intermittently introduced into the lung as tidal volume (VT) by an expiratory valve. ITPV can be combined with pressure control mode, resulting in hybrid ventilation (HV). We hypothesized that HV might decrease VT, compared with volume controlled ventilation (VCV) or pressure controlled ventilation (PCV) alone.
METHODS
HV was compared with VCV and PCV in 7 tracheostomized rabbits. We aimed at maintaining PaCO2 levels normal as the respiratory rates (RR) were set at 20, 40, 80, and 120/min. Blood pressure and airway pressures were monitored and dead space ratio was calculated.
RESULTS
The dead spaces (VD) of VCV are 30+/-4 mL, 18+/-4 mL, 14+/-4 mL, and 12+/-5 mL and the VD of PCV are 24+/-6 mL, 16+/-3 mL, 15+/-4 mL and 12+/-4 mL at the respiratory rates of 20/min, 40/min, 80/min, and 120/min, respectively. The VD of HV are 13+/-6 mL, 9+/-3 mL, 7+/-2 mL, and 5+/-1 mL, respectively. The VT and PIP of HV are significantly lower than those of VCV and PCV at the same RR.
CONCLUSION
It can be concluded that HV can be applied to minimize the airway pressures and dead space ventilation of VCV and PCV.
Key Words: Ventilation; intratracheal pulmonary; hybrid; volume controlled; pressure controlled; Tidal volume; Dead space
TOOLS
Share :
Facebook Twitter Linked In Google+ Line it
METRICS Graph View
  • 2,134 View
  • 5 Download
Related article in NM

New Modes of Ventilation for Neonates.2012 November;19(4)



ABOUT
ARTICLE CATEGORY

Browse all articles >

BROWSE ARTICLES
AUTHOR INFORMATION
Editorial Office
34, Sajik-ro 8–gil(King’s Gargen 3 Block 1207), Jongno-gu, Seoul 03174, Republic of Korea
Tel: +82-2-730-1993    Fax: +82-2-730-1994    E-mail: neonate2002@naver.com                

Copyright © 2024 by The Korean Society of Neonatology.

Developed in M2PI

Close layer
prev next